INTERCEPT Blood System for Platelets
Collection and processing specifications for platelets in PAS-3
Product Detail
Specification
  Small Volume Set (SV) Large Volume Set (LV)
Dual Storage Set (DS)
Platelet source Amicus
Apheresis
Amicus
Apheresis
Amicus
Apheresis
Suspension medium PAS-3 & Plasma (32-47%) PAS-3 & Plasma (32-47%)
PAS-3 & Plasma (32-47%)
Platelet input volume (mL) 255-325 300-390
300-390
375-420
Platelet dose (x1011) 2.9-5.0 3.0-6.0
3.0-6.0
4.0-8.0
Platelet count (x109/mL) 0.9-2.0 0.8-2.0
0.8-2.0
0.95-2.1
RBC (x106/mL) <4 <4
<4
CAD time (hr) 4-16 6-16
6-16
  1 1
1 or 2
Units per carton 24 24 (packaged as 4 foil pouches with 6 processing sets per pouch)

*Platelets must be exposed to UVA light within 24 hours after collection.


Collection and processing specifications for platelets in 100% plasma
Product Detail Specification
  Small Volume Set (SV) Large Volume Set (LV)
Dual Storage Set (DS)
Platelet source Trima
Apheresis
Trima
Apheresis
Trima
Apheresis
Suspension medium 100% Plasma 100% Plasma
100% Plasma
Platelet input volume (mL) 270-325 300-390
300-390
375-420
Platelet dose (x1011) 3.0-5.0 3.0-5.2
3.0-5.2
5.3-7.9
Platelet count (x109/mL) 0.9-2.0 0.9-1.7
0.8-1.7
1.3-2.0
RBC (x106/mL) <4 <4
<4
CAD time (hr) 12-24 12-24
12-24
Integrated storage containers 1 1
1
2
Units per carton 24 24 (packaged as 4 foil pouches with 6 processing sets per pouch)

*Platelets must be exposed to UVA light within 24 hours after collection.


INTERCEPT Blood System for Plasma Collection and processing specifications for plasma
Product Detail Specification
Plasma input volume (mL) 585 - 650
RBC (x106/mL) <4
CAD Flow through
Transfusion units/treatment 2 or 3 (user options)
Transfusion unit volume (mL) ~200-325
Type of collection Apheresis or whole blood*
Integrated storage containers 3
Units per carton 24 (packaged as 4 foil pouches with 6 processing sets per pouch)

*The use of whole blood derived plasma requires pooling of 2-3 units.


INTERCEPT INT100 Illuminator
Product Detail Specification
Compatibility With all INTERCEPT processing sets for platelets (SV, LV, DS) and plasma
Throughput* Platelet: 40 units/hr
Plasma: 36 units/hr
Stackable Yes, up to 2 instruments
Dimensions
(W x D x H)
114.9 x 73.8 x 36.9 cm
(45.2 x 29.1 x 14.5 in)
Weight 68.95 kg (152 lb)
Voltage 120 V (5 A max)
Frequency 60 Hz
Operational temperature 18-30°C
Operational humidity <80%

*Includes preparation, loading and unloading of UVA illuminator. Assumes 2-unit yield from each platelet treatment and 3-unit yield from each plasma treatment.