Product specifications
Procleix Panther System
Sample throughput
- Specification
-
- Standard: 3.5 hr to first result; 5 results every 5 min thereafter
- IDT: 275 results in 8 hr; 500 results in 12 hr
Sample capacity
- Specification
-
120 sample tubes (15 tubes per rack x 8 racks; with continuous sample loading ability)
Sample barcode types
- Specification
-
Code 39, Code 93, Code 128 (ISBT 128), NW7, Interleaved 2 of 5, Codabar, JAN-13, UPC
Reagent capacity
- Specification
-
- 1 reagent kit per lane
- 4 lanes available to load any combination of assay reagents
- 2000 test on-board capacity
- Continuous reagent kit loading ability
Waste capacity
- Specification
-
Process 750 tests before unloading liquid and solid waste
Dimensions
(W x D x H)
- Specification
-
- Instrument: 122 x 81.5 x 175 cm (48 x 32 x 69 in)
- UPS (optional): 21.4 x 41.0 x 32.5 cm (8.4 x 161.1 x 12.8 in)
Note: touchscreen monitor adds 45.5 cm (18 in) to width
Clearance requirements
- Specification
-
- Width: 183 cm, minimum (6 ft; recommended space is 244 cm [8 ft])
- Vertical: 198 cm (6.5 ft)
- Front: 91.4 cm (3 ft)
Note: touchscreen monitor adds 45.5 cm (18 in) to width
Weight
- Specification
-
- Instrument: 363 kg (800 lb)
- UPS: 34.5 kg (76 lb)
Voltage
- Specification
-
- Instrument: 100-240 VAC (±10%; single phase)
- UPS: 110 V-20 amp circuit (dedicated), specified to operate at a minimum of 15 amps (dedicated)
- UPS: 220 V-10 amp circuit (dedicated), specified to operate at 7.5 amps (dedicated)
Frequency
- Specification
-
50/60 Hz, single phase
Power
- Specification
-
Instrument and ancillaries: 1400 W (peak; 13 amps @ 100 VAC, 5.4 amps @ 240 VAC)
Operational temperature
- Specification
-
15-30°C (59-86°F)
Operational humidity
- Specification
-
20-85% (relative, non-condensing)
Operational altitude
- Specification
-
≤2000 m (6562 ft; above sea-level)
Heat dissipation
- Specification
-
550 W (1878 BTU/hr; during steady state)
Drain output
- Specification
-
None; waste fluids drain to removable containers
Noise output
- Specification
-
<70 dBA
Product Detail | Specification |
---|---|
Sample throughput |
|
Sample capacity |
120 sample tubes (15 tubes per rack x 8 racks; with continuous sample loading ability) |
Sample barcode types |
Code 39, Code 93, Code 128 (ISBT 128), NW7, Interleaved 2 of 5, Codabar, JAN-13, UPC |
Reagent capacity |
|
Waste capacity |
Process 750 tests before unloading liquid and solid waste |
Dimensions |
Note: touchscreen monitor adds 45.5 cm (18 in) to width |
Clearance requirements |
Note: touchscreen monitor adds 45.5 cm (18 in) to width |
Weight |
|
Voltage |
|
Frequency |
50/60 Hz, single phase |
Power |
Instrument and ancillaries: 1400 W (peak; 13 amps @ 100 VAC, 5.4 amps @ 240 VAC) |
Operational temperature |
15-30°C (59-86°F) |
Operational humidity |
20-85% (relative, non-condensing) |
Operational altitude |
≤2000 m (6562 ft; above sea-level) |
Heat dissipation |
550 W (1878 BTU/hr; during steady state) |
Drain output |
None; waste fluids drain to removable containers |
Noise output |
<70 dBA |