Technical specifications Physical dimensions (WxDxH) | - Instrument: 122 x 81.5 x 175 cm
- Instrument with MTU Expansion Module: 149 x 81.5 x 175 cm
| * UPS: 21.4 x 41 x 32.5 cm - Direct-to-drain pump: 24 x 22 x 23 cm
|
Minimum clearance requirements | - Base: 183 x 172.9 x 198 cm
- With MTU Expansion Module: 214 x 172.9 x 198 cm
| - No need for rear clearance, instrument can be installed directly against a wall
|
Weight | - Base: 363 kg
- UPS: 34.5 kg
| - With MTU Expansion Module: 423 kg
|
Sample capacity | - Up to 120 sample tubes (8 racks) + on-demand access*
| |
Reagent capacity | - Up to 2000 tests (4 sets) + on-demand access*
| |
Fluids capacity | - Up to 2000 tests (2 sets) + on-demand access*
| |
Consumables capacity | - MTU: Up to 125 tests + on-demand access*
- Tips: Up to 576
| - MTU (with MTU Expansion Module): Up to 600 tests + on-demand access*
|
Liquid waste capacity | - Base model: Up to 750 tests + on-demand access*
| - With optional direct-to-drain: No limit**
|
Solid waste capacity | - Up to 750 tests+ on-demand access*
| |
Sample throughput | - 3.5 hr to first 5 results; 5 results every 5 min thereafter
| - 275 test results in 8 hr; up to 500 test results in 12 hr
|
Sample barcode types | - Code 39, Code 93, Code 128 (ISBT 128), Interleaved 2 of 5, Codabar, Jan 13, UPC
| |
System calibration | | |
On-board reagent stability | - Up to 60 (also non-continuous) hours
| |
Environmental requirements | - Altitude: Equal to or less than 2000 m above sea level
- Operational temperature: 15–30°C
| - Operational humidity: 20–85% relative (non-condensing) at 30°C or below
|
Electrical requirements | - Electrical input: 100–240; VAC: (± 10%), 50/60 Hz: single phase
- Current input: 110 V - 20 amp circuit (dedicated if used with optional UPS) specified to operate at a minimum of 15 amps (dedicated); 220 V - 10 amp circuit (dedicated) specified to operate at 7.5 amps (dedicated)
- Current draw: Draw 1400 W peak, average 700 W (13 amps @ 100 VAC, 5.4 amps @ 240 VAC)
| |
Environmental output | - Heat dissipation: 550 W (1878 BTU/hr) during steady state
- Drain output: Waste fluids drain to removable containers, or directly to centralized laboratory drainage system**
- Noise output: <60 dB(A)
| |
Operating system | | |
MTU Expansion Module (optional) | - Up to 120 MTU capacity (up to 600 tests)
| |
Data sharing (optional) | - Data sharing among up to 16 connected instruments
- Exchange partially utilized reagent kits among connected instruments
- Single user login and setup across networked instruments
- Duplicate-sample prevention across networked instruments
| |
Dashboard (optional) | - Provides real-time status of networked Procleix Panther systems featuring ART that can be displayed via TV screen monitor, tablet or desktop PC***
| |