Procleix Panther System
Product Detail Specification
Sample throughput
  • Standard: 3.5 hr to first result; 5 results every 5 min thereafter
  • IDT: 275 results in 8 hr; 500 results in 12 hr
Sample capacity 120 sample tubes (15 tubes per rack x 8 racks; with continuous sample loading ability)
Sample barcode types Code 39*, Code 93*, Code 128 (ISBT 128), NW7, Interleaved 2 of 5*, Codabar*, JAN-13, UPC
Reagent capacity
  • 1 reagent kit per lane
  • 4 lanes available to load any combination of assay reagents
  • 2000 test on-board capacity
  • Continuous reagent kit loading ability
Waste capacity Process 750 tests before unloading liquid and solid waste
Dimensions
(W x D x H)
  • Instrument: 122 x 81.5 x 175 cm (48 x 32 x 69 in)
  • UPS (optional): 21.4 x 41.0 x 32.5 cm (8.4 x 161.1 x 12.8 in)
Note: touchscreen monitor adds 45.5 cm (18 in) to width
Clearance requirements
  • Width: 183 cm, minimum (6 ft; recommended space is 244 cm [8 ft])
  • Vertical: 198 cm (6.5 ft)
  • Front: 91.4 cm (3 ft)
Note: touchscreen monitor adds 45.5 cm (18 in) to width
Weight
  • Instrument: 363 kg (800 lb)
  • UPS: 34.5 kg (76 lb)
Voltage
  • Instrument: 100-240 VAC (±10%; single phase)
  • UPS: 110 V-20 amp circuit (dedicated), specified to operate at a minimum of 15 amps (dedicated)
  • UPS: 220 V-10 amp circuit (dedicated), specified to operate at 7.5 amps (dedicated)
Frequency 50/60 Hz, single phase
Power Instrument and ancillaries: 1400 W (peak; 13 amps @ 100 VAC, 5.4 amps @ 240 VAC)
Operational temperature 15-30°C (59-86°F)
Operational humidity 20-85% (relative, non-condensing)
Operational Altitude ≤2000 m (6562 ft; above sea-level)
Heat dissipation 550 W (1878 BTU/hr; during steady state)
Drain output None; waste fluids drain to removable containers
Noise output <70 dBA

*Checksum optional