Procleix Tigris System
Product Detail Specification
Sample throughput
  • Standard: 3.5 hr to first result; 109 results every hour thereafter
  • IDT: 500 results in 9 hr
System capacity
  • Continuous on-board capacity: Up to 1000 tests
  • Assay worklist: Up to 500 tests
  • Samples: Up to 180 tubes in 9 sample racks (Note: sample throughput may vary with Procleix assays)
Sample barcode types Code 39*, Code 128 (non-ISBT), ISBT 128, Codabar*, UPC-A, EAN-13, JAN-13, EAN-8, and JAN-8
Sample tube sizes
  • 12 x 75 mm (2.9 in)
  • 13 x 75 mm (2.9 in)
  • 13 x 82 mm (3.2 in)
  • 13 x 100 mm (3.9 in)
  • 16 x 75 mm (2.9 in)
  • 16 x 100 mm (3.9 in)
Waste capacity
  • Solid waste containers: Up to 1000 tests
  • Liquid waste containers: Up to 1000 tests
Dimensions
(W x D x H)
  • Total: 330 x 150 x 240 cm (129.9 x 59.1 x 94.5 in)
  • Analyzer (with standard MTU drawer): 175.3 x 91.4 x 183 cm (69 x 36 x 72 in)
  • Computer workstation: 63.5 x 70.4 x 112 cm (25 x 27.7 x 44 in)
  • UPS: 48.3 x 62.2 x 8.9 cm (19 x 24.5 x 3.5 in)
  • Vacuum pump: 33 x 55.9 x 53.3 cm (13 x 23 x 21 in)
Weight
  • Analyzer (with standard MTU drawer): 585.6 kg (1291 lb)
  • Computer workstation: 36.3 kg (80 lb)
  • UPS: 37 kg (82 lb)
  • Vacuum pump: 35 kg (77 lb)
Voltage 200-240 VAC (16 A)
Frequency 50/60 Hz
Operational temperature 15-25°C (59-77°F)
Operational humidity 20-85%
Operational Altitude 0-2242 m (0-7355 ft)
Heat dissipation 1384 W (4725 BTU/hr; during steady operation)
Noise output Compliant with Standard EN61010-1

*Checksum optional


This information is intended for physicians and healthcare professionals only.